Control beeper

[:IMMediate] <freq, time>
:SYSTem:BEEPer[:IMMediate] <freq, time>
Parameters
freq =
time =
NOTE
The frequency and time values must be separated by a comma (i.e., :syst:beep
100, 3).
Description
The beeper of the SourceMeter can be used to provide an audible signal
at a specified frequency and time duration (up to 7.9 seconds @ 65Hz).
This beeper can, for example, be used to signal the end of a lengthy
sweep.
Example: :SYSTem:BEEPer500, 1 Beep at 500Hz for 1 second
The correlation between the duration and frequency of the beep is
expressed as follows:
Maximum Time = 512/ Frequency
For example, at a frequency of 512Hz, the maximum beep time is one
second. You can set the time greater than one (1) second, but it will only
beep for one second.
Note that in order to use this command, the beeper must be enabled.
(See next command.)
STATe <b>
:SYSTem:BEEPer:STATe <b>
Parameters
<b> =
Query
:STATe?
Description
This command is used to enable or disable the beeper. When enabled, a
short beep is provided to signal that a front panel key has been pressed.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SCPI Command Reference 18-109
65 to 2e6
Specify frequency in Hz
0 to 7.9
Specify time duration
Enable or disable beeper
1 or ON
Enable beeper
0 or OFF
Disable beeper
Query state of beeper