18-100 SCPI Command Reference
DELay <n>
:SOURce[1]:PULSe:DELay <n>
Parameters
<n> =
Query
:DELay?
:DELay? DEFault
:DELay? MINimum
:DELay? MAXimum
Description
This command is used to set the pulse delay for the Model 2430 Pulse
Mode. The pulse delay occurs during the off-time of the pulse period,
and is used to adjust the duty cycle.
NOTE
See
Section 5
SOURce2
The following commands are used to set the logic level of the digital output lines, and con-
trol the pulse width of limit test output patterns that are sent to component handlers. Limit
tests are configured and controlled from the CALCulate2 Subsystem. For details on limit
testing, see
Section
Setting digital output
[:LEVel] <NRf> | <NDN>
:SOURce2:TTL:[LEVel] [:DEFault] <NRf> | <NDN>
:SOURce2:TTL:[LEVel]:ACTual?
Parameters
<NRf> | <NDN> = 0 to 7, 3-bit
Query
:TTL?
:ACTual?
Description
This command is used to set the logic levels of the output lines of the
Digital I/O port. When set high, the specified output line will be at
approximately +5V. When set low, the output line will be at 0V.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Specify pulse delay
0 to 9999.99900 Specify pulse delay in seconds
MINimum
0 seconds
MAXimum
9999.999 seconds
DEFault
0 seconds
Query pulse width
Query *RST default pulse width
Query lowest allowable pulse width
Query highest allowable pulse width
for details on Model 2430 Pulse Mode.
12,
"Limit
Testing."
Set digital output pattern
Query actual output pattern
Specify digital output value
0 to 15, 4-bit
Query default digital output value