I-Source Operating Boundaries - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual

I-Source operating boundaries

Figure 6-9
and
Figure 6-10
quadrant of operation is covered. Operation in the other three quadrants is similar.
2400 —
Figure 6-9A
shown, the SourceMeter can output up to 105mA at 210V, or 1.05A at 21V. Note that
when sourcing more than 105mA, voltage is limited to 21V.
2410 —
Figure 6-9B
shown, the SourceMeter can output up to 21mA at 1100V, or 1.05A at 21V. Note that
when sourcing more than 21mA, voltage is limited to 21V.
2420 —
Figure 6-9C
shown, the SourceMeter can output up to 1.05A at 63V, or 3.15A at 21V. Note that when
sourcing more than 1.05A, voltage is limited to 21V.
2425 and 2430 —
Source. In the DC Mode (solid line), the SourceMeter can output up to 1.05A at 105V, or
3.15A at 21V. Note that when sourcing more than 1.05A, voltage is limited to 21V. The
dotted line in
Figure 6-9D
which is 10.5A at 105V.
2440 —
Figure 6-9E
output up to 1.05A at 42V or 5.25A at 10.5V. When sourcing more than 1.05A, voltage is
limited to 10.5V.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
show the operating boundaries for the I-Source. Only the first
shows the output characteristics for the Model 2400 I-Source. As
shows the output characteristics for the Model 2410 I-Source. As
shows the output characteristics for the Model 2420 I-Source. As
Figure 6-9D
shows the output characteristics of the Model 2430 I-
shows the extended output for the Model 2430 Pulse Mode,
shows Model 2440 I-Source characteristics. The Model 2440 can
Source-Measure Concepts