11-20
Triggering
Output trigger specifications
The SourceMeter can be programmed to output a trigger after various trigger model
actions. (See
"Trigger models," page
compatible output pulse that can be used to trigger other instruments. The specifications
for this trigger pulse are shown in
up to 50mA.
Figure 11-7
Trigger link output pulse specifications
Meter
Complete
TTL High
(3.4V Typical)
TTL Low
(0.25V Typical)
Triggering examples
Front panel trigger example
As an example of how to program the instrument for front panel trigger operation, assume
the trigger parameters below.
model with these operating examples.
Assume the following trigger parameters:
Arm layer source: manual (TRIG key)
Arm layer count: 2
Trigger layer delay: 0.1s
Trigger layer count: 10
Trigger layer output events: source and measure
Trigger layout trigger link output line: #1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
11-30.) The output trigger provides a TTL-
Figure
11-7. A trigger link line can source 1mA and sink
Figure 11-8
shows a flowchart of the front panel trigger
®
User's Manual