Contact Check Programming Example - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

F-16
Contact Check Function

Contact check programming example

Table F-4
summarizes a basic contact check programming example. This example sets up
the SourceMeter as follows:
Contact check: enabled
Contact check threshold resistance: 2Ω
Limit 4 test: enabled
Limit 4 Digital I/O failure pattern: all HI (15)
Contact check event detection: enabled
Contact check timeout: 2 seconds
Measurement function: ohms
Sensing: remote
Table F-4
Contact check programming example
Command
*RST
:SYST:CCH ON
:SYST:CCH:RES 2
:SYST:RSEN ON
:CALC2:LIM4:STAT ON
:CALC2:LIM4:SOUR2 15
:SENS:FUNC "RES"
:TRIG:SEQ2:SOUR CCH
:TRIG:SEQ2:TOUT 2
:OUTP ON
:INIT
:CALC2:LIM4:FAIL?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
Description
Restore GPIB defaults.
Enable contact check.
Set 2Ω contact resistance.
Enable 4-wire remote sense.
Enable contact check pass/fail.
Set Digital I/O bits all HI on failure.
Select ohms function.
Enable contact check event detection.
Set 2 second contact check timeout.
Turn on output.
Trigger unit.
Query contact check pass/fail.
®
User's Manual

Advertisement

Table of Contents
loading

Table of Contents