Ohms Measurements
Figure 4-3
4-wire resistance sensing
SourceMeter
Input/Output HI
4-wire Sense HI
I
V
M
4-wire Sense LO
Input/Output LO

Sense selection

Use the following procedure to select 2-wire or 4-wire sense operation:
Press CONFIG then Ω to display the ohms configuration menu.
1.
2.
Select SENSE MODE, then press ENTER.
3.
Select 2-WIRE or 4-WIRE as desired, then press ENTER.
4.
Press EXIT to return to normal display.