There are two equations for each of the high power ranges (20V, 3A and 100V, 1A). Both
of the equations for the selected voltage range must be true to ensure that the SourceMeter
will not overheat. The result of the first equation must be ≤100, and the result of the sec-
ond equation must be ≤85.
NOTE
For the lower power ranges, the following equations will always be true.
20V, 3A range power equations:
{[(33 × |I
|) - (V
OUT
and
| × 40 × DC)] × 0.35} + T
{[30 + (|I
OUT
100V range power equations:
{[(100 × |I
|) - (V
OUT
and
| × 125 × DC)] × 0.35} + T
{[30 + (|I
OUT
Model 2440 SourceMeter
There are two equations for each of the high power voltage ranges (10V and 40V). Both of
the equations for the selected voltage range must be true to ensure that the SourceMeter
will not overheat. The result of the first equation must be ≤100, and the result of the sec-
ond equation must be ≤85.
10V range power equations:
{[(21 × |I
|) - (V
OUT
and
| × 24 × DC)] × 0.35} + T
{[30 + (|I
OUT
40V range power equations:
{[(58 × |I
|) - (V
OUT
and
| × 120 × DC)] × 0.35} + T
{[30 + (|I
OUT
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Source-Measure Concepts
× I
)] × 0.7 × DC} + T
≤ 100
OUT
OUT
AMB
≤ 85
AMB
× I
)] × 0.7 × DC} + T
≤ 100
OUT
OUT
AMB
≤ 85
AMB
× I
)] × 0.7 × DC} + T
≤ 100
OUT
OUT
AMB
≤ 85
AMB
× I
)] × 0.7 × DC} + T
≤ 100
OUT
OUT
AMB
≤ 85
AMB
6-9