Output Configuration Commands; Remote Output Configuration - Keithley SourceMeter 2400 User Manual

Sourcemeter 2400 series source measure unit
Hide thumbs Also See for SourceMeter 2400:
Table of Contents

Advertisement

®
2400 Series SourceMeter
User's Manual Digital I/O Port, Safety Interlock, and Output Configuration
Remote output configuration
Output configuration commands
Table 13-2
summarizes output configuration commands. These commands include those
to enable and disable the interlock as well as commands to control output off states. See
Section
18,
"OUTPut
Table 13-2
Output configuration commands
Command
:OUTPut:INTerlock:STATe <state>
:OUTPut:INTerlock:TRIPped?
:OUTPut:SMODe <name>
:SOURce:CLEar
:SOURce:CLEar:AUTO <state>
:SOURce:CLEar:AUTO:MODE <name>
1. For the Model 2430 Pulse Mode, the output-off mode is always NORMal. Selecting one of the other output-off states causes error
+831.
2. For the Model 2430 Pulse Mode, auto output-off is always enabled. You can set auto output to be off, but it will not disable until
the DC Mode of operation is selected.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
subsystem" and
"SOURce
subsystem" for more information.
Description
Enable/disable interlock (state = ON or OFF).
Query interlock tripped state (1 = tripped).
Select output-off mode (state = HIMPedance, NORMal, ZERO,
1
or GUARd).
Turn output source off when in idle state.
Enable/disable auto output-off. State = ON (output off after
measurement) or ON (output stays on).
Auto clear mode. Name = ALWays (every reading; default) or
TCOunt (ON when trigger layer entered; OFF when leaving
trigger layer).
13-11