Triggering
NOTE
For the Model 2430 Pulse Mode, an output trigger can only occur after the mea-
sure action. Therefore, the output trigger settings for the source and delay
actions are ignored.
DELAY — Specify the time delay (in seconds) for the trigger delay.
NOTE
For the Model 2430 Pulse Mode, trigger delay is not used. Therefore, the trigger
delay setting is ignored.
COUNT — Specify the trigger count.
HALT — Use to return the SourceMeter to the idle state. HALT does not turn off
the output. The programmed source level will still be available at the OUTPUT ter-
minals. The following actions will take the SourceMeter out of idle:
Turn the output off and then on again.
Re-select the arm or trigger event.
Exit from the menu structure, and then re-enter it by pressing CONFIG and
then TRIG.
Figure 11-2
Trigger configuration menu tree
ARM
LAYER
ARM
ARM IN
COUNT
OUT
LINE
EVENTS
IMMEDIATE
GPIB
TIMER
MANUAL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
2400 Series SourceMeter
CONFIG
TRIG
TRIG
LAYER
TRIGGER
TRIGGER
IN
OUT
TRIGGER
IMMEDIATE
LINE
EVENTS
LINK
↑↓
TLINK
STEST
STEST
STEST
®
User's Manual
HALT
DELAY
COUNT