Remote Programming - Source Output Commands - Keithley 6220 Reference Manual

Ac and dc current source measure unit
Table of Contents

Advertisement

3-16
DC Current Source Operation
(4).
Remote programming – source output commands
Table 3-2
programming example to output DC current is also provided.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Turn on the output
NOTE
In order to turn on the output, an interlock switch
must be connected to the INTERLOCK connector
on the rear panel of the Model 622x. Closing the in-
terlock switch will enable the OUTPUT allowing it
to be turned on. For details, see
page
2-4.
The OUTPUT key toggles the output state (on or off). When the output is
turned on, the OUTPUT indicator light turns on.
For remote programming, the command to control the output is
documented in
Table
3-2. The following example demonstrates proper
syntax.
Example – Turns the output on:
OUTPut ON
The OUTPUT indicator will blink if the current source goes into compliance.
This indicates that the set current is not being delivered to the load. See
"Compliance," on page 3-3
lists the commands to configure and control the DC output. A
Return to
Model 6220/6221 Reference Manual
' Turns output on.
for details.
Section 3 topics
"INTERLOCK," on

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6221

Table of Contents