External triggering

Input and output triggers are received and sent via the rear panel TRIGGER LINK connector.
The trigger link has six lines. At the factory, line #2 is selected for output triggers and line #1 is
selected for input triggers. These input/output assignments can be changed as previously
explained in this section. The connector pinout is shown in Figure 9-4.
Figure 9-4
Trigger link connection
operation

Input trigger requirements

An input trigger is used to satisfy event detection for a trigger model layer that is using the
TLINK control source. The input requires a falling-edge, TTL compatible pulse with the speci-
fications shown in Figure 9-5.
Figure 9-5
Trigger link input
pulse specifications
Rear Panel Pinout
8 7 6
3
5
4
2 1
Trigger Link
Triggers on
Leading Edge
TTL High
(2V-5V)
TTL Low
(<0.8V)
10µs
Minimum
Pin Number
Description
1
Trigger Link 1
2
Trigger Link 2
3
Trigger Link 3
4
Trigger Link 4
5
Trigger Link 5
6
Trigger Link 6
7
Ground
8
Ground