Operation
Instrument Controls Menu
Digital Input Settings
3-48
Model 48i Instruction Manual

Non-Alarm

The Non-Alarm status screen allows the user to select the non-alarm status
for the selected relay output. The selected item is shown by "<--" after it.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > select Relay > Instrument State > Non-Alarm.
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note The digital inputs must be asserted for at least one second for the
action to be activated.
Note Not all of the I/O available in the instrument is brought out on the
supplied terminal board. If more I/O is desired, an alternative means of
connection is required. (See optional
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings.
EXT ALARM 3
O2 CONC MAX
O2 CONC MIN
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE
LOCAL/REMOTE
SERVICE
UNITS
RANGE
AVG
DIAGS ALARM
ZERO MODE
SPAN MODE
SAMPLE MODE
PURGE MODE
DIGITAL INPUT SETTINGS:
>1
NOP
SET BACKGROUND
2
NOP
CAL TO LO SPAN
3
NOP
AOUTS TO ZERO
4
NOP
AOUTS TO FS
5
NOP
RANGE
AVG
DIAGS ALARM
Terminal Block and Cable
NONE
Thermo Fisher Scientific
Kits.)