Table B-14. Analog Voltage Output Range Values - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
analog vout range channel
This command reports the analog voltage output channel range, where
channel is between 1 and 6 inclusive, according to
following example reports that analog voltage output channel 2 is set to 3
(1-10 V).
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range channel range
This command sets analog voltage output channel to the range, where
channel is between 1 and 6 inclusive, and range is set according to
14. The following example sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B–14. Analog Voltage Output Range Values
Range
1
2
3
4
0 [cannot be set to this, but may report]
mb read coils start count
start = index of first coil, count = number of coils to report
This command reports the current state of the MODBUS coils (digital
outputs). Output is in binary format with the coil start appearing as the
right-most bit.
Send:
mb read coils 1 15
Receive:
mb read coils 1 15 000000100000001
mb read registers start count
start = index of first register (must be odd number), count = number of
registers to report (must be even), each pair of registers is reported as a
float.
This command reports the current state of the MODBUS registers (analog
outputs). Output is in floating point format with the pair of registers start
and start+1 appearing as the left-most value.
C-Link Protocol Commands
Table
B–14. The
Output Range
0-1 V
0-100 mV
0-10 V
0-5 V
Undefined
Model 48i Instruction Manual
I/O Configuration
Table B–
B-49