Download Print this page

Thermo Scientific 48i Instruction Manual page 278

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands
Measurement Configuration
B-36
Model 48i Instruction Manual
custom range
range = | 1 | 2 | 3 |
This command reports the user-defined value of custom range 1, 2, or 3.
The following example reports that custom range 1 is defined to 5.50 ppm.
Send:
custom 1
Receive:
custom 1 5.500E+3 ppm
set custom range range value
set custom 1 range value
set custom 2 range value
set custom 3 range value
These commands are used to set the maximum concentration for any of the
three custom ranges 1, 2, or 3 to range value, where value is a floating-point
number representing concentration in ppm or mg/m
example sets the custom 1 range to 55.5 ppm.
Send:
set custom 1 range 55.5
Receive:
set custom 1 range 55.5 ok
range mode
This command reports the current range mode.
Send:
range mode
Receive:
range mode single
set range mode mode
This command sets the current range mode to single, dual, or auto. The
following example sets the range mode to single.
Send:
set range mode single
Receive:
set range mode single ok
gas mode
This command reports the current state of the zero/span valves gas mode as
sample, zero, or span. The following example reports that the gas mode is
sample.
Send:
gas mode
Receive:
gas mode sample
set sample
This command sets the zero/span valves to the sample mode. The following
example sets the instrument to sample mode, that is, the instrument is
reading the sample gas.
3
. The following
Thermo Fisher Scientific

Hide quick links:

Advertisement

loading