Flow - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Diagnostics Menu
Sample/Reference
AGC Intensity
3-64
Model 48i Instruction Manual
PRESSURE:
RANGE

Flow

The Flow screen (read only) displays the flow rate. The flow is measured by
an internal flow sensor. For more information, see Chapter 1,
"Introduction".
In the Main Menu, choose Diagnostics > Flow.
FLOW:
RANGE
The Sample/Reference Ratio screen (read only) displays the ratio of the
intensities of the light source through the sample wavelength and reference
Ratio
wavelength of the bandpass filter wheel. Normally, when zero air is being
sampled, the sample/reference ratio is between 1.14 and 1.18. A ratio
outside may indicate that the filter wheel is dirty or the infrared source is
degraded.
In the Main Menu, choose Diagnostics > Sample/Ref Ratio.
SAMPLE REFERENCE RATIO:
S/R:
RANGE
The AGC Intensity screen (read only) displays the intensity (in Hertz) of
the reference channel Automatic Gain Control (AGC) circuit. The AGC
circuit optimizes the noise and resolution levels of the Model 48i. The
AGC intensity reading should be about 250,000 Hertz.
753.0 mmHg
AVG
DIAGS ALARM
1.108 L/min
AVG
DIAGS ALARM
1.161111
AVG
DIAGS ALARM