Thermo Scientific 48i Instruction Manual page 279

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
Send:
set sample
Receive:
set sample ok
set zero
This command sets the zero/span valves to the zero mode. The following
example sets the instrument to zero mode that is, the instrument is reading
the zero gas.
Send:
set zero
Receive:
set zero ok
set span
This command sets the zero/span valves to the span mode. The following
example sets the instrument to span mode that is, the instrument is reading
span gas.
Send:
set span
Receive:
set span ok
gas unit
This command reports the current gas units (ppm or mg/m
reports that the gas unit is set to ppm.
Send:
gas unit
Receive:
gas unit ppm
set gas unit unit
unit = | ppm | mg/m
3
|
This command sets the gas units to ppm or mg/m
sets the gas units to mg/m
Send:
set gas unit mg/m3
Receive:
set gas unit mg/m3 ok
pres comp
This command reports whether pressure compensation is on or off. The
following example shows that pressure compensation is on.
Send:
pres comp
Receive:
pres comp on
set pres comp onoff
onoff = | on | off |
These commands turn the pressure compensation on or off. The following
example turns pressure compensation off.
Send:
set pres comp off
Receive:
set pres comp off ok
3
.
C-Link Protocol Commands
Measurement Configuration
3
). The example
3
. The following example
Model 48i Instruction Manual
B-37