Download Print this page

Data Treatment; Flag Status Data - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Instrument Controls Menu

Data Treatment

Flag Status Data

3-34
Model 48i Instruction Manual
SET PERCENT LRECS:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
The Data Treatment screen is used to select the data type for the selected
record type: whether the data should be averaged over the interval, the
minimum or maximum measured during the interval, or the current value
(last value measured). Data treatment doesn't apply to all data, just to the
concentration measurement. All other data points log the current value at
the end of the interval.
Note If Data Treatment is set to AVG, Datalogging averaging is done in
addition to the normal concentration averaging.
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Data Treatment.
The Flag Status Data is used to set flag status data On/Off. Refer to "Flags
Status" in Appendix B.
In the Main Menu, choose Instrument Controls > Datalogging Settings
> Configure Datalogging > Flag Status Data.
50%
60% ?
SAVE
SET LREC DATA TYPE:
CURRENTLY:
AVG
SET TO:
CUR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
SET LREC FLAG STATUS DATA:
CURRENTLY: OFF
SET TO: ON

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
SET PERCENT LRECS:
CURRENTLY:
SET TO:
THIS WILL ERASE ALL
PRESS
TO CONFIRM CHANGE
RANGE
AVG
DIAGS ALARM
SAVE
?
SAVE
50%
60% ?
LOGS!x
Thermo Fisher Scientific

Hide quick links:

Advertisement

loading