Operation
Instrument Controls Menu
Data Treatment
Flag Status Data
3-34
Model 48i Instruction Manual
SET PERCENT LRECS:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
AVG
DIAGS ALARM
The Data Treatment screen is used to select the data type for the selected
record type: whether the data should be averaged over the interval, the
minimum or maximum measured during the interval, or the current value
(last value measured). Data treatment doesn't apply to all data, just to the
concentration measurement. All other data points log the current value at
the end of the interval.
Note If Data Treatment is set to AVG, Datalogging averaging is done in
addition to the normal concentration averaging.
In the Main Menu, choose Instrument Controls > Datalogging Settings
●
> Configure Datalogging > Data Treatment.
The Flag Status Data is used to set flag status data On/Off. Refer to "Flags
Status" in Appendix B.
In the Main Menu, choose Instrument Controls > Datalogging Settings
●
> Configure Datalogging > Flag Status Data.