Download Print this page

Voltages - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Diagnostics Menu
Motherboard Voltages
Interface Board Voltages
3-62
Model 48i Instruction Manual

Voltages

The Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating
voltages without having to use a voltage meter. The I/O board is only
displayed if the I/O expansion board option is installed.
In the Main Menu, choose Diagnostics > Voltages.
The Motherboard screen (read only) is used to display the current voltage
readings on the motherboard.
In the Main Menu, choose Diagnostics > Voltages > Motherboard
Voltages.
The Interface Board screen (read only) is used to display the current voltage
readings on the interface board. This menu is only displayed if the I/O
expansion board option is installed.
In the Main Menu, choose Diagnostics > Voltages > Interface Board
Voltages.
PROGRAM VERSIONS:
PRODUCT:
VERSION: 01.06.08.248
FIRMWARE:
RANGE
AVG
DIAGS ALARM
VOLTAGES:
>MOTHERBOARD
INTERFACE BOARD
I/O BOARD
RANGE
AVG
DIAGS ALARM
MOTHERBOARD VOLTAGES:
3.3 SUPPLY
5.0 SUPPLY
15.0 SUPPLY
24.0 SUPPLY
-3.3 SUPPLY
RANGE
AVG
DIAGS ALARM
MODEL 48i
11.08.105
3.3 V
5.0 V
15.0 V
24.1 V
-3.3 V
Thermo Fisher Scientific

Hide quick links:

Advertisement

loading