Allow Over/Under Range - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Operation
Instrument Controls Menu

Allow Over/Under Range

Analog Output Configuration
(Select Action)
3-50
Model 48i Instruction Manual
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config.
OUTPUT CHANNELS:
>ALLOW OVER/UNDER RANGE
ALL VOLTAGE CHANNELS
ALL CURRENT CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
RANGE
AVG
The Allow Over/Under Range screen is used to select whether or not the
analog outputs are allowed to exceed the maximum selected value of 100
mV, 1 V, 5 V, 10 V, or 20 mA or the minimum selected value of 0 V, 0
mA, or 4 mA. By default, this parameter is set to on and 5% over and
under range is allowed for all analog output channels.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Allow Over/Under Range.
ALLOW OVER/UNDER RANGE:
CURRENTLY:
SET TO:
RANGE
AVG
The Analog Output Configuration menu displays a list of the analog
output configuration choices, from which the user selects the parameter to
adjust for the selected output channel. Configuration choices include
selecting range, setting minimum/maximum values, and choosing the
signal to output.
Note The settings are different depending on the AUTO/MANUAL
MODE and RANGE MODE settings. Changing either of these settings
will produce a different set of analog output data items. If either of these
modes is going to be changed on a regular basis, then the user must
configure each set of data separately after changing the modes to each
setting.
DIAGS ALARM
ON
OFF
?
TOGGLE VALUE
DIAGS ALARM
Thermo Fisher Scientific