C-Link Protocol Commands
Datalogging
B-24
Model 48i Instruction Manual
Send:
set lrec per 15
Receive:
set lrec per 15 ok
no of lrec
no of srec
These commands report the number of lrecs and srecs stored memory. The
following example shows that 50 lrecs have been stored in the memory.
Send:
no of lrec
Receive:
no of lrec 50 recs
malloc lrec
malloc srec
These commands report the currently set memory allocation for lrecs and
srecs in percent of total memory.
Send:
malloc lrec
Receive:
malloc lrec 10 %
set malloc lrec value
set malloc srec value
value = 0 to 100
These commands set the percent of memory space allocated for lrecs and
srecs to value, where value is a floating-point number representing percent.
The following example sets the memory allocation for lrecs to 10%.
Note Issuing these commands will clear all the logged data in memory. All
the existing records should be retrieved using appropriate commands, if
required.
Send:
set malloc lrec 10
Receive:
set malloc lrec 10 ok
set copy sp to lrec
set copy sp to srec
set copy sp to stream
These commands copy the current selections in scratch pad (sp) into the
lrec, srec, or streaming data list.
The scratch pad is a temporary memory area which is used to set up lists of
selections for lrec, srec, or streaming data items. The user can copy any of
these lists to the scratch pad, modify individual elements in the list, then
save the scratch pad back to the original list. For more information on how
to edit the scratch pad, see the "sp field" command.
The following example copies the current list in scratch pad into the lrecs
list.
Thermo Fisher Scientific