Select Range - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Select Range

Minimum and Maximum
Thermo Fisher Scientific
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > All Voltage Channels, All Current
Channels, Voltage Channel 1–6 or Current Channel 1–6.
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE
CHOOSE SIGNAL TO OUTPUT
RANGE
The Select Range screen is used to select the hardware range for the selected
analog output channel. Possible ranges for the voltage outputs are: 0-100
mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current outputs are: 0-20
mA and 4-20 mA.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > select Channel > Select Range.
SELECT OUTPUT RANGE:
SELECTED OUTPUT:
CURRENTLY:

RANGE
The Minimum and Maximum Value screens are used to edit the zero (0)
and full-scale (100) values, respectively, in percentages for the selected
Value
analog output channel. See
autorange mode, "HI" or "LO" is displayed to indicate high or low
concentrations. Range Status is visible only in auto range mode. The O
conc items are only displayed if the internal O
The minimum and maximum output value screens function the same way.
The following example shows the set minimum value screen.
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > select Channel > Set Minimum or Maximum
Value.
AVG
DIAGS ALARM
SET TO:
0-100mV ?
CHANGE VALUE
AVG
DIAGS ALARM
Table 3–6
Instrument Controls Menu
V ALL
0-10V
SAVE
for a list of choices. In dual or
sensor option is installed.
2
Model 48i Instruction Manual
Operation
2
3-51