Network Time Protocol
I/O Configuration
Thermo Fisher Scientific
Host Name
The Host Name screen is used to edit the host name. When DHCP is
enabled, this name is reported to the DHCP server.
In the Main Menu, choose Instrument Controls > Communication
●
Settings > TCP/IP Settings > Host Name.
The Network Time Protocol (NTP) Server screen is used to edit the IP
Server
address of the NTP server. An NTP server may be used to periodically
synchronize the instrument's real-time clock with a standard. More
information about the NTP servers and a list of public servers may be
found at http://www.ntp.org.
In the Main Menu, choose Instrument Controls > Communication
●
Settings > TCP/IP Settings > NTP Svr.
The I/O Configuration menu deals with configuration of the analyzer's
I/O system. The analog input configuration is displayed only if the I/O
expansion board option is installed.
DEFAULT GATEWAY:
CURRENT: 192.168.1.15
SET TO: 192.168.1.15
MOVE CURSOR
CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
HOST NAME:
CURRENT: ISERIES
ISERIES
ABCDEFGHIJKLMN
OPQRSTUVWXYZ
0123456789 ./-
RANGE
AVG
DIAGS ALARM
NTP SERVER IP ADDRESS:
CURRENT: 192.168.1.20
SET TO:
192.168.001.015
MOVE CURSOR
CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
Instrument Controls Menu
1
SSSSS
?
BKSP
PAGE
SAVE
Model 48i Instruction Manual
Operation
3-45