Set Custom Ranges; Custom Ranges - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Set Custom Ranges

Custom Ranges

Thermo Fisher Scientific
3
ppm
mg/m
10000
10000
C1
C1
C2
C2
C3
C3
C1, C2, and C3 are custom ranges. For more information about custom
ranges, see
"Set Custom
The Set Custom Ranges menu lists three custom ranges, which are user-
defined. In the ppm or mg/m
3
mg/m
can be specified as a range.
In the Main Menu, choose Range > Set Custom Ranges.
CUSTOM RANGES:
>CUSTOM RANGE 1
CUSTOM RANGE 2
CUSTOM RANGE 3
RANGE
AVG
The Custom Ranges screen is used to define the custom ranges.
The display shows the current custom range. The next line of the display is
used to set the range. To use the custom full-scale range, be sure to select it
(Custom range 1, 2, or 3) in the CO Ranges screen. For more information
about selecting ranges, see
In the Main Menu, choose Range > Set Custom Ranges > Custom
Range 1, 2, or 3.
CUSTOM RANGE 1:
CURRENTLY:
SET TO:

CHANGE VALUE
RANGE
AVG
Ranges" below.
3
mode, any value between 1 and 10000
55.0
75.0
125.0
DIAGS ALARM
"CO
Range" above.
55.5
000055.5

MOVE CURSOR
SAVE
DIAGS ALARM
Operation
Range Menu
?
Model 48i Instruction Manual
3-15