Download Print this page

Thermo Scientific 48i Instruction Manual page 305

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

(0x03/0x04) Read Holding
Registers/Read Input
Registers
Thermo Fisher Scientific
Request
Field Name
Function
Starting Address Hi
Starting Address Lo
Quantity of Outputs Hi
Quantity of Outputs Lo
Response
Field Name
Function
Byte Count
Output Status 2–10
Output Status 11–15
The status of outputs 2–10 is shown as the byte value 0xCD, or binary
1100 1101. Output 10 is the MSB of this byte, and output 2 is the LSB.
By convention, bits within a byte are shown with the MSB to the left and
the LSB to the right. Thus, the outputs in the first byte are '10 through 2',
from left to right. In the last data byte, the status of outputs 15-11 is shown
as the byte value 0x0A, or binary 0000 1010. Output 15 is in the fifth bit
position from the left, and output 11 is the LSB of this byte. The four
remaining high order bits are zero filled.
Read Holding/Input Registers reads the measurement data from the
instrument. Issuing either of these function codes will generate the same
response. These functions read the contents of one or more contiguous
registers.
These registers are 16 bits each and are organized as shown below. All of
the values are reported as 32-bit IEEE standard 754 floating point format.
This uses 2 sequential registers, least significant 16 bits first.
The request specifies the starting register address and the number of
registers. Registers are addressed starting at zero. Therefore, registers
numbered 1–16 are addressed as 0–15. The register data in the response
message are packed as two bytes per register, with the binary contents right
justified within each byte. For each register, the first byte contains the high
order bits and the second contains the low order bits.
(Hex)
0x01
0x00
0x02
0x00
0x0D
(Hex)
0x01
0x03
0xCD
0x0A
MODBUS Protocol
Function Codes
Model 48i Instruction Manual
C-5

Hide quick links:

Advertisement

loading