Thermo Scientific 48i Instruction Manual page 253

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
Averaging Time (seconds)
Selection
7
90
8
120
9
180
10
240
11
300
co
high co
low co
These commands report the measured CO concentration when operating
in single range, or high and low CO when operating in dual or auto range
mode. The following example shows that the CO concentration is 40 ppm.
Send:
CO
Receive:
CO 4.000E+1 ppm
o2
This command reports the oxygen concentration in percent. The following
example reports that the current O
Send:
o2
Receive:
o2 15.00 %
o2 corr conc
This command reports the oxygen correction concentration in percent.
The following example shows that the O
Send:
o2 corr conc
Receive:
o2 corr conc 15.00 %
set o2 corr conc value
value = 0 to 20.9
This command sets the oxygen correction concentration value to value,
where value is a floating-point number representing the oxygen correction
concentration in percent. The following example sets the O
concentration value to 15.00%.
Send:
set o2 corr conc 15
Receive:
set o2 corr conc 15 ok
flow
This command reports the current sample flow. The following example
reports that the current sample flow is 1.108 liters/minute.
C-Link Protocol Commands
concentration is 15%.
2
correction concentration is 15%.
2
Model 48i Instruction Manual
Measurements
corrected
2
B-11