Table B-1. Command Response Error Descriptions; Table B-2. C-Link Protocol Commands - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Entering Units in
Convert Concentration
Formats
Commands List
Thermo Fisher Scientific
Table B–1. Command Response Error Descriptions
Command Response
bad cmd
too high
too low
invalid string
data not valid
can't, wrong settings
can't, mode is service
The "save" and "set save params" commands (duplicated for backward
compatibility) store parameters in FLASH memory. It is important that
this command be sent each time instrument parameters are changed. If
changes are not saved, they will be lost in the event of a power failure.
When interfacing to an instrument via C-link commands, always enter the
concentration values in ppb or μg/m
PPB
background value to 20 ppm, enter 20000 (ppb) as the value for the set
background command.
Convert concentration formats from xxxxE+yy to x.xxxE+yy. The iSeries
uses more standard script than the C-Series by placing only one digit to the
left of the decimal.
Table B–2
lists the 48i C-Link protocol commands. The interface will
respond to the command strings outlined below.
Table B–2. C-Link Protocol Commands
Command
1
2
3
4
addr dns
addr gw
addr ip
Description
Command is not recognized
Supplied value is higher than the upper limit
Supplied value is lower than the lower limit
Supplied string invalid (typically because a letter was detected
when the value should be numeric)
Supplied value is not acceptable for entered command
Command not allowed for current measurement mode
Command not allowed while instrument is in service mode
3
units. For example, to set a
Description
Simulates pressing soft key 1 pushbutton
Simulates pressing soft key 2 pushbutton
Simulates pressing soft key 3 pushbutton
Simulates pressing soft key 4 pushbutton
Reports/sets domain name server address for Ethernet port
Reports/sets default gateway address for Ethernet port
Reports/sets IP address for Ethernet port
C-Link Protocol Commands
Commands
Page
B-33
B-33
B-33
B-33
B-40
B-41
B-41
Model 48i Instruction Manual
B-3