set alarm chamber temp min value
set alarm chamber temp max value
These commands set the bench temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing bench temperature alarm limits in degrees C. The following
example sets the bench temperature alarm maximum value to 55.0 °C.
Send:
set alarm chamber temp max 55.0
Receive:
set alarm chamber temp max 55.0 ok
alarm conc min
alarm conc max
These commands report the CO concentration alarm minimum and
maximum values current setting. The following example reports that the
CO concentration minimum is 5.2 ppm.
Send:
alarm conc min
Receive:
alarm conc min 5.2 ppm
set alarm conc min value
set alarm conc max value
These commands set the CO concentration alarm minimum and
maximum values to value, where value is a floating-point representation of
the concentration alarm limits. Values must be in the units that are
currently set for use. The following example sets the CO concentration
alarm maximum value to 215.
Send:
set alarm conc max 215
Receive:
set alarm conc max 215 ok
alarm conc o2 min
alarm conc o2 max
These commands report the current oxygen concentration alarm minimum
and maximum value settings. The following example reports that the O
concentration alarm minimum value is 10%.
Send:
alarm conc o2 min
Receive:
alarm conc o2 min 10.00 %
set alarm conc o2 min value
set alarm conc o2 max value
These commands set the oxygen concentration alarm minimum and
maximum values to value, where value is a floating-point number
representing the oxygen concentration alarm limits in percent. The
following example sets the O
100.00%.
concentration alarm maximum value to
2
Model 48i Instruction Manual
C-Link Protocol Commands
Alarms