Hide thumbs Also See for 48i:

Advertisement

Flow Transducer
Calibration
Thermo Fisher Scientific
4. Remove the two retaining screws holding the flow transducer to the
floor plate and remove the flow transducer
Figure 7–12. Replacing the Flow Transducer
5. Install the new flow transducer following the previous steps in reverse.
6. Calibrate the flow transducer. Refer to the
Calibration" procedure that follows.
Use the following procedure to calibrate the flow transducer.
Equipment Required:
Calibrated flow sensor
WARNING The service procedures in this manual are restricted to qualified
service representatives.
If the equipment is operated in a manner not specified by the
manufacturer, the protection provided by the equipment may be
impaired.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
1. Remove the instrument cover.
2. Disconnect the pump cable from AC PUMP connector on the
measurement interface board.
Retaining Screws (2)
Servicing

Flow Transducer Calibration

(Figure
7–12).
"Flow Transducer
Model 48i Instruction Manual
7-25