Operation
Instrument Controls Menu
Default Gateway
3-44
Model 48i Instruction Manual

IP Address

The IP Address screen is used to edit the IP address. The IP address can
only be changed when DHCP is off. If DHCP is on, the instrument will
respond with "NOT SETTABLE IF DHCP IS ON". For more
information on DHCP, see
In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > IP Address.
Netmask
The Netmask screen is used to edit the netmask. The netmask is used to
determine the subnet on which the instrument can directly communicate
to other devices. The netmask can only be changed when DHCP is off. If
DHCP is on, the instrument will respond with "NOT SETTABLE IF
DHCP IS ON". For more information on DHCP, see
above.
In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > Netmask.
The Default Gateway screen is used to edit the gateway address. The
default gateway can only be changed when DHCP is off. If DHCP is on,
the instrument will respond with "NOT SETTABLE IF DHCP IS ON".
For more information on DHCP, see
addresses that are not on the local subnet will be routed through this
address.
In the Main Menu, choose Instrument Controls > Communication
Settings > TCP/IP Settings > Gateway.
"Use
IP ADDRESS:
CURRENT: 192.168.1.200
SET TO: 192.168.1.200 1

MOVE CURSOR

CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
NETMASK:
CURRENT: 255.255.255.0
SET TO: 255.255.255.0 0

MOVE CURSOR

CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
DHCP" above.
"Use
"Use
DHCP" above. Any traffic to
DHCP"