Test Analog Outputs - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Test Analog Outputs

Set Analog Outputs
Instrument
Configuration
Thermo Fisher Scientific
RELAY STATE:
>OUTPUT 1
OUTPUT 2
OUTPUT 3
OUTPUT 4
OUTPUT 5
RANGE
AVG
The Test Analog Outputs menu is used to set each of the analog output
channels to zero or full-scale. Channel choices include all analog outputs, 6
voltage channels, and 6 current channels (if the I/O expansion board
option is installed).
In the Main Menu, choose Diagnostics > Test Analog Outputs.
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
RANGE
AVG
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-
scale voltage, zero sets the analog outputs to 0 volts, and reset returns the
output to normal operation. The analog outputs are returned to normal
operation upon exiting this screen. The following example shows the screen
when all analog outputs are set to "normal" operating mode.
In the Main Menu, choose Diagnostics > Test Analog Outputs > ALL,
Voltage Channel 1–6, or Current Channel 1–6.
SET ANALOG OUTPUTS:
SETTING:
OUTPUT SET TO:
 SET TO FULL SCALE
 SET TO ZERO
 RESET TO NORMAL
RANGE
AVG
The Instrument Configuration screen displays information on the
hardware configuration of the instrument.
0
0
0
1
1
DIAGS ALARM
DIAGS ALARM
ALL
NORMAL
DIAGS ALARM
Model 48i Instruction Manual
Operation
Diagnostics Menu
3-67