RS-485 Connection

Ethernet Connection
External Accessory
Connector
Thermo Fisher Scientific
The instrument uses a four wire RS-485 configuration with automatic flow
control (SD). Refer to
Table 8–2. RS-485 DB9 Connector Pin Configuration
DB9 Pin
Function
2
+ receive
8
- receive
7
+ transmit
3
- transmit
5
ground
An RJ45 connector is used for the 10Mbs Ethernet connection supporting
TCP/IP communications via standard IPV4 addressing. The IP address
may be configured for static addressing or dynamic addressing (set using a
DHCP server).
Any serial port protocols may be accessed over Ethernet in addition to the
serial port. Up to three simultaneous connections are allowed per protocol.
The external accessory connector is not used in the Model 48i analyzer.
This port is used in other models to communicate with smart external
devices that may be mounted hundreds of feet from the analyzer using an
RS-485 electrical interface.
Table 8–2
for the DB9 connector pin configuration.