Table B-10. Allow Mode Command Values - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

C-Link Protocol Commands
Communications Configuration
B-46
Model 48i Instruction Manual
upon, but it will not change the instrument lock status (this is for
compatibility with systems expecting and "ok" response).
Note The instrument will always respond to the command "mode" with
the status of the password lock as "mode local" or "mode remote",
regardless of the above setting.
The following example sets the instrument to accept the "set mode local"
and "set mode remote" commands.
Send:
set allow mode cmd 1
Receive:
set allow mode cmd 1 ok
Table B–10. Allow Mode Command Values
Allow Mode Command
Value
0
Ignore (default)
1
Accept
power up mode
This command reports the current power up mode setting to either 0 =
local/unlocked or 1 = remote/locked, according to
value is 0; power up in local/unlocked mode. The following example shows
that the instrument is configured to power up in the remote/locked mode.
Send:
power up mode
Receive:
power up mode 1
set power up mode value
This command is used to configure the instrument to power up in the
local/unlocked mode (value = 0) or the remote/locked mode (value = 1),
according to
Table
B–11.
If the instrument is set to power up in the local/remote mode, the keypad
can be used to make changes via the front panel. If the instrument is set to
power up in the remote/locked mode, changes can not be made from the
front panel. The following example sets the instrument to power up in
remote/locked mode.
Send:
set power up mode 1
Receive:
set power up mode 1 ok
Table
B–11. The default
Thermo Fisher Scientific