alarm pressure min
Receive:
alarm pressure min 650 mmHg
set alarm pressure min value
set alarm pressure max value
These commands set the pressure alarm minimum and maximum values to
value, where value is a floating-point number representing pressure alarm
limits in millimeters of mercury. The following example sets the pressure
alarm maximum value to 790 mmHg.
Send:
set alarm pressure max 790
Receive:
set alarm pressure max 790 ok
alarm sample flow min
alarm sample flow max
These commands report the sample flow alarm minimum and maximum
value current settings. The following example reports that the sample flow
alarm minimum value is 0.350 L/min.
Send:
alarm sample flow min
Receive:
alarm sample flow min 0.350 l/min
set alarm sample flow min value
set alarm sample flow max value
These commands set the sample flow alarm minimum and maximum
values to value, where value is a floating-point number representing sample
flow alarm limits in liters per minute. The following example sets the
sample flow alarm maximum value to 1 L/min.
Send:
set alarm sample flow max 1.000
Receive:
set alarm sample flow max 1.000 ok
alarm trig conc
This command reports the CO concentration alarm trigger action for
minimum alarm, current setting, to either floor or ceiling. The following
example shows the CO concentration minimum alarm trigger to ceiling,
according to
Table
B–4.
Send:
alarm trig conc
Receive:
alarm trig conc 1
set alarm trig conc value
This command sets the CO concentration alarm minimum value, where
value is set to either floor or ceiling, according to
example sets the CO concentration minimum alarm trigger to ceiling.
Send:
set alarm trig conc 1
Receive:
set alarm trig conc 1 ok
C-Link Protocol Commands
Alarms
Table
B–4. The following
Model 48i Instruction Manual
B-17