Thermo Scientific 48i Instruction Manual page 273

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Thermo Fisher Scientific
span dev
This command reports the span deviation (span check offset). The
following example reports that the span deviation is 1 ppb.
Send:
span dev
Receive:
span dev 1.000 E+00
set span dev value
This command sets the span deviation (span check offset) to value, where
value is a floating-point representation of the gas concentration in current
selected units. The following example sets the span deviation to 345 ppb.
Send:
set span dev 345
Receive:
set span dev 345 ok
span dur
This command reports the span duration. The following example reports
that the span duration min is 10 minutes.
Send:
span dur
Receive:
span dur 10 min
set span dur value
This command sets span duration to value where value represents the span
duration in minutes. The following example sets the span duration to 15
minutes.
Send:
set span dur 15
Receive:
set span dur 15 ok
zero cal reset
This command reports that the zero cal reset is on or off. The following
example reports that the zero cal reset is off.
Send:
zero cal reset
Receive:
zero cal reset off
set zero cal reset onoff
onoff = | on | off |
This command sets the zero cal reset on or off. The following example
turns the zero cal reset off.
Send:
set zero cal reset off
Receive:
set zero cal reset off ok
C-Link Protocol Commands
Calibration
Model 48i Instruction Manual
B-31