Table 1-2. Model 48I Optional Internal Oxygen Sensor Specifications - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Introduction
Specifications
1-4
Model 48i Instruction Manual
Operating temperature
Power requirements
Physical dimensions
Weight
Analog outputs
Digital outputs
Digital inputs
Serial Ports
Ethernet connection
*In non-condensing environments. Performance specifications based on operation with 20–30 °C range.
Table 1–2. Model 48i Optional Internal Oxygen Sensor Specifications
Technology
Range
Accuracy (Intrinsic error)
Linearity
Repeatability
Zero Drift
Response Time (T
-T
)
10
90
Weight
20–
C (may be safely operated over the range of 0–
30 °
100 VAC @ 50/60 Hz
115 VAC @ 50/60 Hz
220–240 VAC @ 50/60 Hz
275 watts
16.75" (W) X 8.62" (H) X 23" (D)
Approximately 49 lbs.
6 voltage outputs; 0–100 mV, 1, 5, 10 V (user selectable), 5% of
full-scale over/under range (user selectable), 12 bit resolution,
measurement output user selectable per channel
1 power fail relay Form C, 10 digital relays Form A, user
selectable alarm output, relay logic, 100 mA @ 200 VDC
16 digital inputs, user select programmable, TTL level, pulled
high
1 RS-232 or RS-485 with two connectors, baud rate 1200–
115200, data bits, parity, and stop bits, protocols: C-Link,
MODBUS, Gesytec (Bayern-Hessen), and streaming data (all
user selectable)
RJ45 connector for 10Mbs Ethernet connection, static or
dynamic TCP/IP addressing
Paramagnetic (Pm)
0-100% O
2
<±0.1% O
2
<±0.1% O
2
<±0.1% O
2
<±0.2% O
per month (excludes up to 0.1% O
2
hours of operation)
<2.5 seconds
Approximately 2 lbs. (in addition to standard instrument)