Download Print this page

Units - Thermo Scientific 48i Instruction Manual

Gas filter correlation co analyzer
Hide thumbs Also See for 48i:

Advertisement

Decimal Places
Number of Table Points
Thermo Fisher Scientific
ANALOG INPUT DESCRIPTOR:
CURRENTLY: IN1
RANGE

Units

The Units screen allows the user to enter the units for the selected analog
input channel. The units are displayed on the diagnostic screen and in
datalogging and streaming data. The units may be from 1 to 3 characters in
length, and defaults to V (volts).
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Units.
ANALOG INPUT UNITS:
CURRENTLY: V
RANGE
The Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal, from 0 to 6, with a default of 2.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Decimal Places.
DECIMAL PLACES:
CURRENTLY:
RANGE
The Number of Table Points screen allows the user to select how many
points are used in the analog input conversion table for the selected
channel. The instrument uses linear interpolation between the points in
this table to determine what the reading value is based on the analog input
voltage. Each point in the table consists of an analog input voltage value
(0–10.5 V) and a corresponding reading value. Only two points are
IN1
ABCDEFGHIJKLMN
OPQRSTUVWXYZ
0123456789 ./-
AVG
DIAGS ALARM
VN1
ABCDEFGHIJKLMN
OPQRSTUVWXYZ
0123456789 ./-
AVG
DIAGS ALARM
SET TO:

CHANGE VALUE
SAVE VALUE
AVG
DIAGS ALARM
Instrument Controls Menu
BKSP
PAGE
SAVE
BKSP
PAGE
SAVE
2
3 ?
Model 48i Instruction Manual
Operation
3-55

Hide quick links:

Advertisement

loading