Operation
Instrument Controls Menu
3-56
Model 48i Instruction Manual
necessary for linear inputs, however a larger number of points may be used
to approximate non-linear inputs. The points range from 2 to 10, with a
default of 2.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > Table Points.

Table Point

The Table Point submenu allows the user to set up an individual table
point.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel > Point 1-10.
Volts
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default table is
a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U, where U is the previously entered unit of measure.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > select Channel > select Point > Volts.
NUMBER OF TABLE POINTS:
CURRENTLY:
SET TO:

CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
TABLE POINT 01 CONFIG:
>VOLTS
USER VALUE
RANGE
AVG
DIAGS ALARM
TABLE POINT 01 VOLTS:
CURRENTLY:
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
2
10 ?
0.00
0.00
0.00
00.00
SAVE