Motor Speed

O
Sensor Readings
2
Analog Input
Thermo Fisher Scientific
In the Main Menu, choose Diagnostics > AGC Intensity.
The Motor Speed screen (read only) displays the status of the chopper
motor. A reading of 100.0% means that the motor speed is correct. A
reading other than 100.0% indicates that there is a problem with the
chopper motor or power supply.
In the Main Menu, choose Diagnostics > Motor Speed.
The O
concentration (in percent) and the oxygen sensor temperature (in degrees
C). This screen is only visible if the internal O
In the Main Menu, choose > Diagnostics > O2 Sensor Readings.
The Analog Input Readings screen (read only) displays the 8 current user-
scaled analog readings (if the I/O expansion board option is installed).
Readings
In the Main Menu, choose Diagnostics > Analog Input Readings.
AGC INTENSITY:
INTENSITY:
RANGE
AVG
DIAGS ALARM
MOTOR SPEED:
SPEED:
RANGE
AVG
DIAGS ALARM
Sensor Readings screen (read only) displays the measured oxygen
2
O2 CONC:
TEMP:
RANGE
AVG
DIAGS ALARM
250000 Hz
100.00 %
sensor option is installed.
2
15.02 %
o
30.2
C
Model 48i Instruction Manual
Operation
Diagnostics Menu
3-65