MFP OPTIONS: B246/D052 SERIES
12. Make sure the ROM number and firmware version on the diagnostic report are
the same as those in the chart below:
[A]: "ROM Number/Firmware Version" – "HDD Format Option"
[B]: "Loading Program" – "GW1a_zoffy" (
"Loading Program" – "GW2a_zoffy" (
Diagnostic Report:
Data Overwrite Security
Unit Type C
Data Overwrite Security
Unit Type D
Important: The ROM number and firmware version number change when the
firmware is upgraded.
• If the ROM numbers are not the same, or the version numbers are not the same,
this means the unit was not installed correctly.
If this happens:
Make sure the MFP model and unit type match (Type C or D).
The B246 series uses a Data Overwrite Security Unit Type C.
If they do not match:
1) Do the installation procedure again, from Step 1.
2) Replace the "Data Overwrite Security Unit" (SD card) with the correct
type. Redo installation procedure.
3) Replace the NV-RAM. Redo installation procedure.
13. Turn "Auto Erase Memory Setting" ON:
User Tools >> System Settings >> Administrator Tools >> Auto Erase Memory
Setting >> ON
If the customer has enabled Administrator Authentication Management -
Machine Management they will have to log in to change this User Tool
setting.
14. Exit User Tools.
B064 Series/B140 Series
/B246 Series/D052 Series
CÓPIA NÃO CONTROLADA
"ROM No. / Firmware
Version" [A]
HDD Format Option:
B7355050 / 0.04
HDD Format Option:
B7355060 / 0.03
1-190
CÓPIA NÃO CONTROLADA
Data Overwrite Security Unit Type C)
Data Overwrite Security Unit Type D)
"Loading Program" [B]
GW1a_zoffy:
B7355050 / 0.04
GW2a_zoffy:
B7355060 / 0.03
Rev. 09/2006
SM