TR Menu -> DC Control Voltage -> Auto Freq Control -> Tolerance
USER Menu -> DC Control Voltage -> Auto Freq Control -> Tolerance
:SOURce:VOLTage:CONTrol:CORRection:COLLect:ACQ
uire
Syntax

:SOURce:VOLTage:CONTrol:CORRection:COLLect:ACQuire

Description
Executes DC CTRL DRIFT CAL (No Query)
Equivalent key
No equivalent key is available on the front panel.

:SOURce:VOLTage:CONTrol:CORRection[:STATe]

Syntax
:SOURce:VOLTage:CONTrol:CORRection[:STATe] {ON|OFF|1|0}
:SOURce:VOLTage:CONTrol:CORRection[:STATe]?
Description
Sets/Reads DC CTRL DRIFT CAL state
Parameter
ON or 1
OFF or 0(Preset
value)
Equivalent key
No equivalent key is available on the front panel.

:SOURce:VOLTage:CONTrol:DELay

Syntax
:SOURce:VOLTage:CONTrol:DELay <numeric>
:SOURce:VOLTage:CONTrol:DELay?
Description
Sets/Reads DC Control delay setting(sec)
Parameter
Range
Preset value
Unit
Resolution
Equivalent key
PN Menu -> DC Control Voltage -> DC Control Delay
7
:SOURce:VOLTage:CONTrol:CORRection:COLLect:ACQuire
Description
Set DC CTRL DRIFT CAL state mode to 'ON'
Set DC CTRL DRIFT CAL state mode to 'OFF'
<Numeric>
0 to 1
100m
s
1m
SCPI Command Reference