Agilent Technologies 8753ES Datasheet

Agilent Technologies 8753ES Datasheet

Network analyzers, 30 khz to 3 or 6 ghz
Hide thumbs Also See for 8753ES:

Advertisement

This document describes the performance
and features of the following products:
• Agilent 8753ES S-parameter vector network
analyzer
• Agilent 8753ES Option 011 vector network
analyzer without test set
• Agilent 8753ET Transmission/reflection vector
network analyzer
Agilent 8753ET/ES
8753ET and 8753ES Network
Analyzers, 30 kHz to 3 or 6 GHz
Data Sheet
For more information about these analyzers,
please see the following documents:
Agilent 8753ET and 8753ES
Network Analyzers
Overview
Agilent 8753ET and 8753ES
Network Analyzers
Configuration Guide
Pub. Number
5968-5159E
5968-5158E

Advertisement

loading

Summary of Contents for Agilent Technologies 8753ES

  • Page 1 This document describes the performance and features of the following products: • Agilent 8753ES S-parameter vector network analyzer • Agilent 8753ES Option 011 vector network analyzer without test set • Agilent 8753ET Transmission/reflection vector network analyzer Agilent 8753ET/ES 8753ET and 8753ES Network...
  • Page 2: Table Of Contents

    Table of Contents Definitions and Test Conditions System Performance Summaries Specifications and Characteristics Measurement Throughput Summary Options Test Set Specifications for Option 011 Accessories...
  • Page 3: Definitions And Test Conditions

    The measurement uncertainty curves and meas- urement port characteristics given for Agilent Technologies 8753ES systems also apply to the 8753ES with Options 006 and 011 and the 85047A test set (50-ohm), or the 8753ES Option 011 with an 85046B test set (75-ohm).
  • Page 4: System Performance Summaries

    System Performance Summaries Agilent 8753ES (50-ohm systems) 7-mm test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85031B Test-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 50 kHz to 300 kHz 90 dB...
  • Page 5 Agilent 8753ES (50-ohm systems) Type-N test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85032B Test-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 50 kHz to 300 kHz 90 dB 300 kHz to 16 MHz...
  • Page 6 System Performance Summaries Agilent 8753ES (50-ohm systems) 3.5-mm test ports Configuration Network analyzer 8753ES Standard and Option 006 Calibration kit 85033D Test-port cables 11857D System dynamic range 30 kHz to 50 kHz 70 dB 50 kHz to 300 kHz 90 dB...
  • Page 7 Agilent 8753ES (75-Ohm Systems) Type-N test ports Configuration Network analyzer 8753ES Option 075 Calibration kit 85036B Test-port cables 11857B System Dynamic Range 30 kHz to 50 kHz 68 dB 50 kHz to 300 kHz 90 dB 300 kHz to 16 MHz...
  • Page 8 System Performance Summaries Agilent 8753ES (75-ohm systems) Type-F test ports Configuration Network analyzer 8753E Option 075 Calibration kit 85039B Test-port cables 11857B System dynamic range 30 kHz to 50 kHz 68 dB 50 kHz to 300 kHz 90 dB 300 kHz to 16 MHz...
  • Page 9 Agilent 8753ET (50-ohm systems) Type-N test ports Configuration Network analyzer 8753ET Standard, Option 006, or Option 004 Calibration kit 85032B Test port cable Part number 8120-4781 System dynamic range 300 kHz to 16 MHz 100 dB 16 MHz to 3 GHz 110 dB 3 GHz to 6 GHz 105 dB...
  • Page 10 9. Typical performance. 10. 25° ±5° C. 11. The Agilent 8753ES source characteristics and measurement accuracy in this mode are dependent on the stability of the external LO source. The RF source tracks the LO to maintain a stable IF signal at the R-channel receiver input.
  • Page 11 <0.006 dB rms 3 GHz to 6 GHz <0.010 dB rms 1. See the Agilent 8753ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal. 2. Typical performance. 3. Marker resolution for magnitude; phase and delay is dependent upon measured value.
  • Page 12 3. 16 MHz to 3 GHz. 4. 16 MHz to 2 GHz. 5. For Agilent 8753ES Option 011 and Option 006, linearity is specified for the ranges of -5 to +13 dBm and +13 to +18 dBm. 6. Typical performance.
  • Page 13 Degradation in accuracy is negligible when using an Agilent 8642A/B, 8656B, or E4432B RF signal generator as the LO source. 4. See the Agilent 8753ES descriptions and options for a functional description. Measurement accuracy is dependent on the stability of the input signal.
  • Page 14 Agilent 8753ET Specifications Test-Port Output Characteristics Frequency characteristics Range 300 kHz to 3 GHz Option 006 300 kHz to 6 GHz Resolution 1 Hz Stability ±7.5 ppm (0° to 55° C) ±3 ppm/year Option 1D5 ±0.05 ppm (0° to 55° C) ±0.5 ppm/year Accuracy ±10 ppm at 25°...
  • Page 15 External source mode (CW time sweep only) Frequency range 300 kHz to 6 GHz R-channel input requirements Power level 0 to -25 dBm Spectral purity Maximum spurious input <-30 dBc Residual FM <20 kHz Typical settling time 500 ms (automatic) 50 ms (manual) Frequency readout accuracy 0.1% (automatic)
  • Page 16 Measurement parameters 8753ET: Reflection, transmission, A, B, R, A/R, B/R, A/B. Conversion to impedance or admittance. 8753ES: S11, S21, S12, S22, A, B, R, A/R, B/R, A/B. Conversion to impedance or admittance. Formats Cartesian: log/linear magnitude, phase, group delay, SWR, real and imaginary.
  • Page 17 Enters the velocity factor to calculate equivalent electrical length. Reference plane extension Redefine the plane-of-measurement reference to other than port 1 or port 2 of the 8753ET and 8753ES. Select default calibration kit Select from a list of standard calibration kits or choose a user-defined kit.
  • Page 18 Agilent 8753ET/ES Supplemental Characteristics (continued) Trace Functions Display data Display current measurement data, memory data, or current measurement with measurement and memory data simultaneously. Trace math Vector division or subtraction of current linear measurement values and memory data. Display annotations Start/stop, center/span, or CW frequency, source level, scale/div, reference level, marker data, softkey functions, warning and caution messages,...
  • Page 19 Frequency offset on/off Sets the RF source to be swept at a fixed offset frequency above the receiver as required in a swept RF/IF, fixed LO, mixer test. Service menu Select the desired service test, service diagnostic, service or verification mode. Test Sequences Description Create, edit, save, or recall a series of front-panel key-...
  • Page 20: General Characteristics

    This mini-DIN connector is used for adding an IBM PC-AT compatible keyboard. Test set interconnect This connector is used to connect an 8753ES Option 011 to the 85046A/B or 85047A test set. 8753ES analyzers without Option 011 can use signal levels on this connector for sequencing or general purpose I/O applications.
  • Page 21: Measurement Throughput Summary

    Environmental Characteristics General Conditions RFI and EMI susceptibility: defined by VDE 0730, CISPR Publication 11, and FCC Class B Standards. ESD (electrostatic discharge): must be eliminated by use of static-safe work procedures and an anti- static bench mat. The flexible rubber keypad pro- tects key contacts from dust, but the environment should be as dust-free as possible for optimal relia- bility.
  • Page 22: Test Set Specifications For Option 011

    The test sets are controlled from the 8753ES Option 011 and include a programmable step attenuator. The frequency range of the 85046A 50-ohm test set is 300 kHz to 3 GHz. The 85046A has precision 7-mm connectors.
  • Page 23 3 MHz to 6 GHz in a single sweep or switched out to measure 300 kHz to 3 GHz in a single sweep. The 8753ES Option 011 does not use the frequency doubler, so the full 300 kHz to 6 GHz range is available.
  • Page 24: Accessories

    DC to 6 GHz 85033D 3.5-mm calibration kit Contains a set of precision 3.5-mm standards to calibrate the 8753ES and 50-ohm test sets for the measurement of devices with precision 3.5-mm and SMA connectors. Precision phase-matched 7-mm to 3.5-mm adapters are included for accurate measurements of non-insertable devices.
  • Page 25 3.5-mm female to 7-16 female 1 adapter 85039B type-F calibration kit Contains a set of 75-ohm type-F standards to cali- brate the 8753ES with a 75-ohm test set for the measurement of devices with type-F connectors. Description F-male 75-ohm termination...
  • Page 26 ±2° at 1.3 GHz 11857D 7-mm test-port return cables A pair of 610-mm (24-in.) test-port return cables for use with the 8753ES with the 85046A and 85047A S-parameter test sets. These cables can be used with connector types other than 7-mm with the appropriate precision adapters.
  • Page 27 The 11853A 50-ohm type-N and the 11854A 50-ohm BNC accessory kits provide the RF components generally required when using either the 85046A, 85047A, or 11850C with the 8753ES Option 011 when measuring devices having 50-ohm type-N or BNC connectors. These kits are supplied with a storage case.
  • Page 28 U.S. registered trademark of Microsoft Corporation. Agilent Technologies’ Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need.