Disassembly Guide

Replacement

Power Entry
Module (PEM)

Removal/

Replacement
Removal
56
5. Connect the leads to the battery. The red wire connects to the positive
terminal, and the black wire connects to the negative terminal.
6. Insert the new battery into the bottom case with the negative terminal
towards the outside of the monitor. Install the bracket and grounding lead
with the two screws.
7. Complete the procedure in paragraph Monitor Assembly on page 54.
8. Turn the monitor on and verify proper operation.
1. Follow the procedure in paragraphs Prior to Disassembly on page 52, and
Monitor Disassembly on page 53.
2. Push the top of the Power Entry Module (PEM) in from the outside of the
case, and lift up.
3. Use needle-nose pliers to disconnect the leads from the PEM (see
Figure 11 on page 57).