Performance Verification

Operational Setup

Setting the Time and Date

22
The operational setup procedure sets the time and date into the monitor. Refer to
the O
M
N-65 Operator's Manual for individual parameter settings. The
XI
AX
monitor returns to the factory default settings when it is powered on: therefore,
individual parameter settings need to be set at that time.

Discussion

WARNING: The O
M
XI
O
M
N-65 when the patient alarm events are recorded to the memory
XI
AX
chip in the sensor. The accuracy of the date/time used for patient alarm
events depends on the date/time set in the monitor.
Note: When the monitor times-out in 30 seconds, the parameter is set and the
monitor display returns to the normal operation.
Normal operation means:
the monitor is turned on
a sensor is connected to the monitor
the sensor is applied to the patient
the patient's %SpO
readings (BPM) are being reported
no error conditions exist

Procedure

To set the time and date:
1. With the monitor in normal operation, press the Power button to turn the
monitor off.
2. Press the Power button to turn the monitor on.
sensor uses the date and time provided by the
AX
(oxygen saturation percentage) and pulse rate
2
O
M
N-65 Service Manual
XI