Operational Setup

N-595
23. Press the LIMITS softkey. Verify that the %SpO
boxed and indicates an alarm limit of "100."
24. Verify that the %SpO
alarm limit of "85."
25. Verify that the SatSeconds SAT-S alarm is set to OFF.
26. Verify that the pulse rate upper alarm limit display is boxed and indicates an
alarm limit of "170."
27. Verify that the pulse rate lower alarm limit display is boxed and indicates an
alarm limit of "40."
28. Press the ON/STANDBY button to turn the monitor off.
Operational setup procedures verify and set up the following parameters.
Alarms and Alarm Silence on page 18
Alarm Volume Control on page 20
Pulse Tone Volume Control on page 20
Nurse Call on page 21
Analog Output on page 22
Operation on Battery Power on page 23
lower alarm limit display is boxed and indicates an
2
Performance Verification
upper alarm limit display is