Table 9: Operational Performance Problems - Nellcor OxiMax N-550 Service Manual

Hide thumbs Also See for OxiMax N-550:
Table of Contents

Advertisement

Troubleshooting
Operational Performance
48
Table 8: Display/Alarms Problems (Continued)
Symptom
An alarm condition
exists but no alarm
(audible or visual) is
indicated.
Table 9 lists symptoms of problems relating to operational performance (no
error codes displayed) and recommended actions. If the action requires
replacement of a PCB or module, refer to Disassembly Guide on page 55.

Table 9: Operational Performance Problems

Symptom
The pulse amplitude
indicator seems to
indicate a pulse, but the
digital displays show
zeroes.
SpO
or pulse values
2
change rapidly; Pulse
Amplitude indicator is
erratic.
Recommended Action
Replace the Front Panel PCB. Refer to Front Case
Disassembly on page 58.
Recommended Action
The sensor may be damaged; replace it.
Replace the Front Panel PCB. Refer to Front Case
Disassembly on page 58.
The sensor may be damp or may have been reused too
many times. Replace it.
An electrosurgical unit (ESU) may be
interfering with performance:
Move the N-550 and its cables and sensors as
far from the ESU as possible.
Plug the N-550 power supply and the ESU
into different AC circuits.
Move the ESU ground pad as close to the
surgical site as possible and as far away from
the sensor as possible.
Verify the performance with the procedures
detailed in Performance Verification on
page 7.
Replace the Front Panel PCB. Refer to Front
Case Disassembly on page 58.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents