Performance Test and Function Test
Performance Test

DC Bias Level Accuracy Test (Option 001)

This test verifies the E4991A's DC bias level accuracy at several level.
Specification
Voltage(open terminal)
Current(short terminal)
Required Equipment
Description
Multimeter
BNC(m)-BNC(m) Cable, 61 cm
N(m)-BNC(f) Adapter
7mm-N(f) Adapter
Dual Banana-BNC(f) Adapter
Procedure
1. Connect the test equipment as shown in Figure 2-11. The BNC(f)-Banana adapter must
be connected to voltage measurement connectors with the "GND" connector
connecting to the "LO" terminal.
42
(0.1 % + 6 mV)
(0.2 % + 20 A)
Recommended Model
3458A
p/n 8120-1839
p/n 1250-0780
11524A
p/n 1251-2277
Chapter 2