Power Level
Frequency
30 dBm
10 MHz
35 dBm
2 GHz
40 dBm
100 MHz
Chapter 2
Test Limit
Test Result
2.00 dB
3.00 dB
2.00 dB
Performance Test and Function Test
Test Record
Measurement
Uncertainty
0.22 dB
dB
0.24 dB
dB
0.27 dB
dB
TUR
9.38
12.69
7.63
75