How To Update The E4991A Firmware - Agilent Technologies E4991A Service Manual

Rf impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Figure C-2
Properties Dialog Box (example)
5. Perform 1 through 4 for each disk.
NOTE
If the label name is not assigned, the firmware update for the E4991A will not be done
correctly.

How to update the E4991A firmware

It is possible to update the E4991A firmware, but not the following software by this
procedure.
OS (Operating System)
VBA (Visual Basic for Application)
Calibration Constant Data
The following is the procedure to update the E4991A firmware.
Step 1. Prepare the E4991A firmware installation disks (three disks).
Step 2. Connect the mouse and the keyboard to the E4991A rear panel. Then turn the E4991A on.
Step 3. Exit from the E4991A Measurement Display.
1. Click
appear.
2. Enter the password
3. Click the
Step 4. Double-click the "My Computer" icon on the Windows desktop.
Step 5. Insert the E4991A firmware installation disk1 (1 of 3) into the FDD of the E4991A.
Step 6. Double-Click the "3 1/2 Floppy [A:]" icon.
Appendix C
-
on the menu bar, and the Enter Password to exit dialog box will
System
Exit
into the Password box.
e4991a
button. Then windows desktop screen is displayed.
OK
Firmware Update
Firmware Update
335