Adjustment
A2 DC Bias Voltage Adjustment (Opt 001 only)
A2 DC Bias Voltage Adjustment (Opt 001 only)
The purpose of this procedure is to adjust the output voltage and the voltage monitor on the
A2 DC Bias Board.
Required Equipment
Multimeter
7mm-N(f) Adapter
N(m)-BNC(f) Adapter
BNC(f)-Banana Adapter
BNC Cable, 61 cm
Procedure
Step 1. Run the adjustment program.
Step 2. Choose the A2 Adjustment(Opt.001) in the Spot Adjustment.
It is possible to perform this item with the All Adjustment too.
Step 3. Connect the instrument as shown in Figure 3-18.
Figure 3-18
A2 DC Bias Voltage Adjustment Setup
Step 4. Follow the adjustment program instruction to update the correction constant.
144
Description
3458A
11524A
p/n 1250-0780
p/n 1251-2277
p/n 8120-1839