Performance Test and Function Test
Performance Test
Figure 2-6
Segment Table 1(Power Level: -3 dBm, Averaging: 8)
6. Press
7. Press
8. Press
9. Select
10. Press
11. Select
12. Select
13. Connect the OPEN Termination included in the 16195B to the 7 mm connector using
the Torque Wrench(p/n:8710-1760).
14. Select
15. Connect the SHORT Termination included in the 16195B to the 7 mm connector using
the Torque Wrench(p/n:8710-1760).
16. Select
17. Connect the 50  Termination included in the 16195B to the 7 mm connector using the
Torque Wrench(p/n:8710-1760).
18. Select
19. Select
20. Press
21. Select
22. Press
23. Set Osc[dBm] to 20 and Avg to 1 for all measurement points as Figure 2-7.
36
.
[Sweep]
-Segment to use segment table.
Sweep Type
.
[Trigger Setup]
-Manual.
Trigger Source
to show the Cal/Compen menu.
[Cal/Compen]
.
Cal Menu[Uncal]
User Freq & Pwr.
Cal Type-
to calibrate open termination.
Meas Open
to calibrate short termination.
Meas Short
to calibrate 50  termination.
Meas Load
.
Done
to show the Save/Recall softkey menu.
[Save/Recall]
.
Save State
, then E4991A's status is saved to "-3.sta" status file.
[-],[3],[Enter]
Chapter 2